RIGOL Chapter 2 Basic Waveform Output

DG4000 Series User’s Guide

2-6

Unit Switch

Vpp is the unit for signal peak-peak value and Vrms is the unit for signal
effective value. The default unit is Vpp. Users can quickly switch the current
amplitude unit from the front panel.

For different waveforms, the relations between Vpp and Vrms are different.
The relation of the two units is as shown in the figure below (take sine
waveform as an example).

Vpp=2Vamp

Vrms=0.707Vamp

Vamp

According to the figure above, the conversion relation between Vpp and
Vrms fulfills the following equation:

Vrms

2

2

Vpp

=

For example, to convert 2Vpp to the corresponding value in Vrms,
press · in the numeric keyboard and select Vrms menu. For sine
waveform, the converted value is 707.2mVrms.