To create new arbitrary waveform, To create new arbitrary waveform -10 – RIGOL DG4000 Series User Manual
Page 86
RIGOL Chapter 3 Arbitrary Waveform Output
DG4000 Series User’s Guide
3-10
To Create New Arbitrary Waveform
Press Arb, use
to open the 2/2 menu page, and then press Create New to
open the arbitrary waveform creation interface.
1. Cycle Period
Press this softkey and use the numeric keyboard or direction keys and knob to
set the cycle period of the new waveform. Cycle period is the boundary time of
the waveform and the range is from 25.0ns to 1Ms.
Note that the time of the last point that can be defined should be lower than
the cycle period.
2. High VLimit
Press this softkey and use the numeric keyboard or direction keys and knob to
set the level upper limit of the new waveform. The level upper limit is the
maximum voltage that can be set during waveform edit. The level upper limit
must be greater than the “Low VLimit” currently set and lower than or equal
to +5V (50Ω).
3. Low VLimit
Press this softkey and use the numeric keyboard or direction keys and knob to
set the level lower limit of the new waveform. The level lower limit is the
minimum voltage that can be set during waveform edit. The level lower limit
must be greater than or equal to -5V (50Ω) and lower than the “High VLimit”
currently set.
4. Init #Points
When creating a new waveform, the waveform editor will automatically create
a waveform with two points. By default, point 1 is located at 0 seconds and
point 2 at half of the specified cycle period. In addition, the levels of the two
points are equal to the specified “Low VLimit”. Press this softkey and use the
numeric keyboard or direction keys and knob to set the initial points of the new
waveform. DG4000 allows users to create arbitrary waveform with 1 to 16384
(16K) points. If in need, you can “Insert Point” or “Delete Point” later.