13-9
Power Voltage
100V to 240V (45Hz to 440Hz)
Power Consumption
Less than 50W
Fuse
250V, T2A
Display
Type
7-inch TFT LCD
Resolution
800 Horizontal × RGB × 480 Vertical Resolution
Color
16M color
Environment
Temperature Range
Operating: 10℃ to 40℃
Non-Operating: -20℃ to 60℃
Cooling Method
Cooling by fans compulsively
Humidity Range
Less than 35℃: ≤90% Relative Humidity (RH)
35℃ to 40℃: ≤60% Relative Humidity (RH)
Altitude
Operating: Less than 3000 meters
Non-Operating: Less than 15000 meters
Mechanical
Dimensions (W×H×D)
313 mm ×160.7 mm×116.7mm
Weight
without package: 3.2 kg
with package: 4.5 kg
Interfaces
USB Host, USB Device, LAN
IP Protection
IP2X
Calibration Interval
Recommend 1 year for standard interval