Chapter 12 Troubleshooting

RIGOL

DG4000 Series User’s Guide

12-1

Chapter 12 Troubleshooting


This chapter lists the commonly encountered failures of DG4000 and their
solutions. When you encounter those problems, please solve them following the
corresponding steps. If the problem persists, please contact

RIGOL and provide

your device information (Utility SystemSys Info).

1. The screen is still dark (no display) after power on:

(1) Check if the power is correctly connected.
(2) Check if the power switch is really on.
(3) Restart the instrument after finishing the above inspections.
(4) If it does not work correctly, contact

RIGOL for our service.

2. The settings are correct but no waveform is generated

(1) Check if the BNC cable is correctly connected to the channel output

terminal ([Output1] or [Output2]).

(2) Check if the BNC cable can work correctly.
(3) Check the Output1 or Output2 button, if it is turned on.
(4) Set PowerOn as “Last” and then restart the instrument after finishing

the above inspections.

(5) If it does not work correctly, contact

RIGOL for our service.


3. The USB storage device cannot be recognized:

(1) Check if the USB storage device can work normally.
(2) Make sure the USB storage device is USB flash storage device. The

generator doesn’t support hard drive-based USB storage device.

(3) Restart the instrument, reinsert the U-disk and check it.
(4) If the USB storage device still can not be recognized, please contact

RIGOL.