Single-Ended and Differential Connections to Analog Inputs

In Personal Daq applications, thermocouples should not be connected single-ended. Doing
so can result in noise and false readings. This is especially true when acquiring other high-
amplitude signals in conjunction with thermocouple signals that are connected single-ended
.

Measurement Duration Column

⎯ Indicates the amount of time used for sampling a channel’s input

signal. The measurement durations range from very slow (610 milliseconds/sample) to very fast
(12.5 milliseconds/sample). Some related terms are defined as follows:

Sample rate – Samples per second. The sample rate is the number of samples that take place per second.
With a slow measurement duration of 610 milliseconds, there will only be 1.6 samples per second. With a
very fast measurement duration of 12.5 milliseconds, there will be 80 samples per second.

Resolution (Bit RMS) – The number of reliable data bits that exist for a signal’s measurement. The greater
the resolution, the more detailed the reading, for example, with increased resolution a reading of 5.12 V
could become 5.11896 V. Personal Daq actually provides for 24 bits of data information; however, the
accuracy of the least significant bits becomes less as the measurement duration speeds up. At a
measurement duration of 610 milliseconds, the last two bits are considered unreliable, resulting in a
resolution of 22 bits. At a very fast measurement duration (12.5 milliseconds), the nine least significant
bits
are unreliable, resulting in 15 bit accuracy.

Speed vs. Resolution

1

Speed Designation

Measurement

Duration

(Scan Period)

Maximum

Sample Rate

2

(Samples/sec)

Resolution

(Bits RMS)

(-4 V to +4 V range)

Very Slow, 50 / 60 Hz rejection

610 ms

1.6 / sec

22

Slow, 50 Hz rejection

370 ms

2.7 / sec

22

Slow, 60 Hz rejection

310 ms

3.2 / sec

22

Medium, 50 Hz rejection

130 ms

7.7

/ sec

21

Medium, 60 Hz rejection

110 ms

9.2 / sec

21

Medium

40 ms

25 / sec

19

Fast

20 ms

48 / sec

17

Very Fast

12.5 ms

80 / sec

15

Notes:

1. Each channel can have independent measurement speed and resolution.

2. The sample rates shown were obtained with continuous self-calibration disabled.

The table values do not include the use of CJC measurements.