23 Power Failure Recovery

The power failure recovery is used to re-create and restart the test that was
running before the power failure. A test that was created and not running
before the power failure will be re-created but not started. The
configuration of the test is saved automatically once the test is created. The
logger, injections, and configuration are periodically saved.

If the power failure occurred while the test case is created, the saved
configuration will be loaded when the IQS-8500 Series and
IQS-8120NGE/8130NGE is manually started from IQS Manager. Thus, the
test that was running will be re-created, configured, and restarted.

Note: For RFC 2544, the configuration will be re-loaded but the test will not

restart automatically.

Note: The power failure recovery is disabled when the SUI terminates normally

or when the test case is cleared.

When Using the Test Timer

Refer to Timer Configuration on page 126 for more information on test
timer.

The test that was running will be re-created and started after a power
failure if all the following conditions are met:

The test was running.

The start time has not expired during the power failure.

The stop time or the duration has not expired during the power failure.

When Using SmartMode

SmartMode is not supported meaning that SmartMode will return to its
default factory settings after a power failure recovery.