beautypg.com

EXFO IQS-8500 Series for IQS-600 User Manual

Page 158

background image

Port Tabs

144

FTB-8500 Series and FTB-8120NGE/8130NGE

Interface Setup (Ethernet)

Flow Control

Available with Frame Analyzer (IQS-8510B and IQS-8510G), RFC 2544,
and TCP Throughput stest only. Flow control is not supported (set to
None) for EtherSAM, BERT and Smart Loopback tests. When Flow
Control
is enabled, the Ethernet and Fibre Channel Application will
stop transmitting for the requested time value specified by the valid
flow control frame received. Choices are None, Enable RX, Enable
TX
, and Enable RX and TX. Only None and Enable RX are available on
IQS-8510G. Enable TX and Enable RX are only available when
Auto-Negotiation is disabled.

None (IQS-8510B/IQS-8120NGE/IQS-8130NGE/IQS-8525/IQS-8535):
pause frame received are ignored and no pause frame are transmitted.

Note: Flow Control should be set to None with Half Duplex mode.

Local Clock Only available for 1Gbps electrical when auto-negotiation
is disabled. Choices are Local, Remote, and Automatic. The default
setting is Automatic.

This manual is related to the following products: