Auto Tests

60

AXS-200/635i

Running Final/Active Pair Test and Viewing Results

Result 1

The Result 1 tab allows you to view the
results of the measured values and
pass/fail status for the Final/Active Pair
tests.

Result 2

The Result 2 tab displays more results
from the Final/Active Pair tests.

³

Met. Insulation displays the
metallic insulation values for the
circuit, ranging from 0 to
999 MOhm.

³

Opens Meter measures the
capacitance of the loop in nF, for
each lead combination.

³

Opens Length displays the corresponding capacitance in either meter
or feet, depending on the unit selection in Regional Settings. If the
corresponding capacitance is < 0.2 nF or > 10 μF, N/A is displayed.

³

Opens Variance is the percentage difference between the capacitance
on T-G and R-G.

³

Load Coil displays In Progress while running and the number of load
coils detected when the test is completed. If the number is 1 or 2,
Probable Filter Detected is displayed. Splitter Detected is displayed if
present.

Press the Details button to switch to the Load Coil tab.