0 appendix, 1 error messages, Appendix – Rice Lake Static Weighing User Manual
Page 64

60
CW-90/90X Checkweigher
9.0
Appendix
9.1
Error Messages
If an error code appears on the display, use the information in Table 9-1 as a troubleshooting guide. If the error
cannot be cleared, call RLWS Service for assistance.
Table 9-1. CW-90/CW-90X Error Messages
Error Display
Description
Solution
- - - - - -
Over range
•Check load cell wiring, including sense jumpers.
•Check configuration, including number of grads
and display divisions.
•Check calibration, including W ZERO and W
SPAN values.
•Under range can be caused by OIML setting and
weight less than -20 display divisions
•Check for scale binding or damage.
•Check for proper excitation voltage.
•Call RLWS Service for assistance.
- - - - - -
Under range
- - - - - -
- - - - - -
A/D out of range
CHKERR
Checksum error
•on power up CHKERR will appear
briefly, then BATTERY BACK LOST
will scroll
•on a failure to write to NVRAM
Press ENTER after replacing the battery to restore
defaults to battery backed storage.
CFGERR
Configuration error on power up if
there was an error loading
configuration
Press ENTER to reboot the indicator. Contact
RLWS Service if problem reoccurs.
ERROR
Internal program error
Check configuration. Run XE command (see
Section 9.2 on page 61) to determine error type.
Call RLWS Service if unable to clear error by
cycling power or if error reoccurs.
HWFERR
Hardware failure error
• on failure to write to the EEPROM
• any error when exiting the menu
Reboot the unit. Call RLWS Service if unable to
clear error by cycling power or if error reoccurs.
LOWBATT
Low battery error flashes every 30
seconds when the battery is low.
Replace the battery.
NOLOAD
ID not loaded
The IDs over/under/target/or tare do not match
the current count by.
FAILED
Failed to acquire target value
Attempt to acquire the current weight on the scale
to the current ID failed.
NO ID
The requested ID was not found
Verify that you are entering the ID correctly and
that the ID exists.
BUFFER
The buffer is nearing capacity when
using the buffer or WeightVault
functionality.
•If using Buffer - dump the buffer contents to a PC
by using the DUMPBUFFER command. After
successfully receiving the buffer contents on the
PC send the CLEARBUFFER command to empty
the buffer.
•If using WeighVault - establish a connection to
the WeighVault PC so that it can retrieve
transactions from the CW-90.
FULL
The buffer is full when using the buffer
or WeightVault functionality.
INVALID UNITS
ID uses unassigned units
Modify the ID to use a configured unit.
NOTARE
Tare is prevented
Change regulatory mode settings or the TAREFN
parameter.
RANGE
GRADS > 100,000
WVAL > 999999.0
Configuration value out of range
A value entered in configuration is out of the
acceptable range.
Ranges for all parameters are listed in the
Parameter Settings Commands section.