beautypg.com

10 appendix, 1 technical data, Appendix – Metrohm 833 IC Liquid Handling Unit User Manual

Page 99: Technical data, Sect. 10 appendix

background image

10.1 Technical data

830 IC Interface / Gebrauchsanweisung 8.830.1001

91

10 Appendix

10.1 Technical

data

Unless anything to the contrary is mentioned, the

published data represent typical values for the 833 IC

Liquid Handling Unit at an ambient temperature of 25°C.

10.1.1 Basic instrument: 833 IC Liquid Handling Unit 1.833.0010

Pump

Pump type

2-channel peristaltic pump with

20

min

-1

(50 Hz)

24

min

-1

(60 Hz)

Capacity

(with water, no

counterpressure)

Typically

with at:

pump tubing

20 min

-1

24

min

-1

6.1826.010

1.41 mL/min 1.69 mL/min

6.1826.030

0.40 mL/min 0.48 mL/min

6.1826.040

0.75 mL/min 0.90 mL/min

6.1826.050

0.43 mL/min 0.52 mL/min

6.1826.070

2.55 mL/min 3.06 mL/min

Pressure

max. 4 bar (0.4 MPa)

Pumpable liquids

Clear liquids with no solid particles

Pump tubing material

PVC (Tygon

ST), PVC (Tygon

LFL), PP

Mains connection

Voltage

115 V: 100...120 V

±

10 %

230 V: 220...240 V

±

10 %

Switched with voltage selection insert in fuse

holder (see Section 2.2.1)

Frequency

50...60 Hz

Power consumption

50 VA

Fuses

5 mm dia., 20 mm long

100

120 V: 0.5 A (slow-blow)

220