833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

52

delivering the sample. One channel of each pump is used to transfer

solution to the dialysis cell, the second channel is required to remove

solution from the dialysis cell. This setup ensures constant flow and

equal pressure conditions on both sides of the dialysis membrane.

7.2 Electrical

connection

7.2.1 Dialysis

without suppression

The electrical connections of the system, consisting of 819 IC Detector,

820 IC Separation Center (2.820.0X10; 1-channel), 818 IC Pump, 833 IC

Liquid Handling Dialysis Unit, 833 IC Liquid Handling Pump Unit and

830 IC Interface are shown in Fig. 16:

Fig. 16:

Connecting the 833 IC Liquid Handling Dialysis Unit to a modular
dialysis system without suppression

Make sure that the remote control of the peristaltic pumps

833 Pump

and

833 Dialysis via 6.2143.210 cable (see Section 3.1.2)

corresponds to the remote configuration of the detector program (see

Section 7.3.1).

830

820

818

819

PC

6.2134.100

6.2134.090

6.2134.130

6.2143.210

6.2115.070

6.2125.090

6.2134.040

833 Pump

833 Dialysis