3.3 Operation
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
25
Start pump with startup hardware
Automatic start of pump drive with
Startup
hardware
.
Time program
A user-specific time program can be entered on the tab
Program
in the
833 IC Liquid Handling Unit
window. Depending on the settings made in
the
Start mode
window (for details see Section 4.4.3 8.110.8281
Instructions for Use for the Metrodata «IC Net 2.3» software), this
program is started automatically either at the start of the determination
(
Start with determination
) or when the sample is injected (
Start with inject
).
Time (1st column)
Time at which program instruction is applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is
applied together with the last instruction with
time entry.
Command (2nd. column)
Program instruction (see below).
Parameter (3rd column)
Parameter for program instruction (see below).
ENABLED
Enable program start (a disabled program is
not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).