5 changing the filter, Changing the filter, Fig. 25 – Metrohm 833 IC Liquid Handling Unit User Manual
Page 89: Filter unit peek 6.2821.120, Ion 9.2.5)
9.2 Care and maintenance
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
81
3 Set the contact pressure
•
Press contact lever 5 until the solution is just aspirated. Then
press the contact lever up by 1 further click to achieve the
optimum contact pressure.
9.2.5
The 6.2821.120 Filter unit PEEK 22 (see Fig. 25) serves to avoid
contamination by abrasive particles of the pump tubings. The filter unit
consists of the housing 82, the filter 83 and the connector 84 to be
screwed into the housing 82. For the connection of capillaries PEEK
compression fittings 17 (6.2744.010) must be used.
New filter 83 are available as an option with the ordering number
6.2821.130 (10 pieces).
To change the filter 83 dismantle the Filter unit PEEK 22 according to
Fig. 25, replace the filter 83 and put it together again. Install the Filter
unit PEEK 22 according to Fig. 7 in the flow path again.
For the connection of the filter unit, please note the flow direction
arrow printed on the housing.
The filter unit is filled with isopropanol when new. Rinse your IC system
carefully after the first installation of a new filter unit.
Fig. 25:
Filter unit PEEK 6.2821.120
16 6.1803.0X0 Aspiration
tube/capillary
PTFE tubing/capillary, depends on
version, see Section 10.2 Standard
17 6.2744.010 PEEK compression
delivery side for connection to:
suppressor
module
sample preparation module
dialysis
cell
ultrafiltration
cell
82 Filter housing
83 Filter 6.2821.130
84 Filter connector
17
16 17
84
83
82 23