9.2 Care and maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

81

3 Set the contact pressure

Press contact lever 5 until the solution is just aspirated. Then

press the contact lever up by 1 further click to achieve the

optimum contact pressure.

9.2.5

Changing the filter

The 6.2821.120 Filter unit PEEK 22 (see Fig. 25) serves to avoid

contamination by abrasive particles of the pump tubings. The filter unit

consists of the housing 82, the filter 83 and the connector 84 to be

screwed into the housing 82. For the connection of capillaries PEEK

compression fittings 17 (6.2744.010) must be used.
New filter 83 are available as an option with the ordering number

6.2821.130 (10 pieces).
To change the filter 83 dismantle the Filter unit PEEK 22 according to

Fig. 25, replace the filter 83 and put it together again. Install the Filter

unit PEEK 22 according to Fig. 7 in the flow path again.

For the connection of the filter unit, please note the flow direction

arrow printed on the housing.
The filter unit is filled with isopropanol when new. Rinse your IC system

carefully after the first installation of a new filter unit.

Fig. 25:

Filter unit PEEK 6.2821.120

16 6.1803.0X0 Aspiration

tube/capillary

PTFE tubing/capillary, depends on

version, see Section 10.2 Standard

equipment

17 6.2744.010 PEEK compression

fitting

23 Discharge capillary

delivery side for connection to:

suppressor

module

sample preparation module

dialysis

cell

ultrafiltration

cell

82 Filter housing

83 Filter 6.2821.130

84 Filter connector

17

16 17

84

83

82 23