3 connecting to 830 ic interface, Connecting to 830 ic interface, Fig. 6 – Metrohm 833 IC Liquid Handling Unit User Manual
Page 26
3 Basic instrument
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
18
3.1.3
Connecting to 830 IC Interface
Connect one of the event lines of the 830 IC Interface with one of the
required remote connections 14 and 15 on the 833 IC Liquid Handling
Unit. The peristaltic pump of the Liquid Handling Unit is controlled via
remote connection Remote Pump 15, this must always be connected.
The connection Remote Actuator 14 is only required for the 833 IC
Liquid Handling Suppressor Unit and the 833 IC Liquid Handling
Sample Preparation Unit; it is used to switch the suppressor or
sample preparation module. The particular modules are connected
to the socket Module 13.
Fig. 6:
Connecting the 833 IC Liquid Handling Unit to 830 IC Interface
For each remote connection one 6.2128.180 Cable is required. At the
830 IC Interface 2
×
7 remote connections can be assembled. Which
event lines are used for this does not matter; when a system is set up
the assignment must only be given correctly in the «IC Net» software
when creating a system, see Section 6.1 of the «IC Net» Instructions for
Use.
816
833
6.2128.180
816
833
Cable
to
Module
830
6.2128.180
830
6.2128.180