beautypg.com

Metrohm 833 IC Liquid Handling Unit User Manual

Page 85

background image

9.1 Faults and their remedies

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

77

Fault

Cause

Remedy

Inadequate or no supply
at all from pump

Dialysis cell

Dialysis cell inlets blocked

Sample tubing blocked by too
large particles

Bacterial growth in sample
channel and dialysis cell

Narrowed PTFE tubing by
compression fittings tightened
too much

Filtration cell

Filter membrane blocked,
filtration cell inlets blocked

Narrowed PTFE tubing by
compression fittings tightened
too much

Replace dialysis membrane and clean
the cell (see Section 9.2.7.1).

Replace blocked tubing, always
centrifuge sample (see Section 9.2.4).

Replace tubing16, 57 und 58 for sample
channel regularly, replace dialysis
membrane

Use capillary cutter to cut off narrowed
tubing ends or replace tubing.


Replace filter membrane and clean the
cell (see Section 9.2.8.1).

Use capillary cutter to cut off narrowed
tubing ends or replace tubing.

Leaks

Tubing nipple leaks

Tighten up or replace tubing nipples.

Air bubbles in pump
circulation system

Aspiration tubing not immersed

Tubing nipple leaks

Dialysis cell

Acceptor not properly
degassed

Standing air bubbles

Completely immerse aspiration tubing.

Tighten up or replace tubing nipples.

Dialysis cell

Degas acceptor solution again.

Small air bubbles do not interfere. For
larger air bubbles rinse the sample
channel with degassed acceptor solution
(possibly open outlet briefly).

Reduced phosphate
sensitivity and/or large
baseline increase
(suppressor)

Suppressor capacity reduced
by heavy metal load or organic
contaminants

Regenerate suppressor (see Section
9.2.6.1
) or replace it (see Section 9.2.6.3).

Suppressor module /
Sample preparation
module
switches to next
position, but does not
work properly

Suppressor module/Sample
preparation module dry

Actuator rotor temporarily in
wrong position






Actuator step not set correctly

Rinse all 3 actuator units with solution for
at least 10 min.

Switch instrument off and on again.
Activate suppressor with . Check
that actuator rotor is in correct position
(see Section 9.2.6.3).
Repeat process 2-3 times, before each
switch to next position with wait
at least 10 s.
If necessary set actuator step again (see
below).

Readjust actuator step
(see Section 9.2.6.3).