beautypg.com

2 fuses, 3 mains cable, 4 switching the instrument on/off – Metrohm 833 IC Liquid Handling Unit User Manual

Page 21: Fuses, Mains cable, Switching the instrument on/off, Fig. 3, Setting the mains voltage, Ction 2.2.2, Fig. 3 )

background image

2.2 Mains connection

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

13

Fig. 3:

Setting the mains voltage

2.2.2 Fuses

The fuse holder of the 833 IC Liquid Handling Unit contains one of the

two fuses 0.5 A/slow blow for 100

120 V or 0.25 A/slow blow for

220

240 V.

Make sure that the instrument is never operated with a different type of

fuse as otherwise it represents a fire hazard!

Checking and replacing a fuse is described in Section 2.2.1.

2.2.3 Mains

cable

The instrument is supplied with one of the following mains cables:

6.2122.020 with SEV 12 plug (Switzerland,

)

6.2122.040 with CEE(7), VII plug (Germany,

)

6.2133.070 with NEMA 5-15 (USA,

)

which has 3 wires and is fitted with a plug with a grounding pin. If a

different plug has to be attached then the yellow/green wire (IEC

standard) must be connection to the grounding pin.

Any break in the grounding inside or outside the instrument will make

it dangerous!

Insert the mains cable plug in mains connection 9 of the 833 IC Liquid

Handling Unit (see Fig. 2).

2.2.4

Switching the instrument on/off

The 833 IC Liquid Handling Unit is switched on and off with mains

switch 8 (see Fig. 2). When the instrument is switched on mains lamp 1

lights up.

100 – 120 V

220 – 240 V

8 Mains switch

9 Mains

connection

11 Fuse holder

220

- 240 V

100 - 120

V

220 - 240

V

100

- 120 V

8

9

11