Metrohm 833 IC Liquid Handling Unit User Manual
Page 36
4 833 IC Liquid Handling Pump Unit
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
28
1 Connect column to injector
•
Remove end caps from column 53.
•
without column heating:
Screw inlet end of separating column 53
(note flow direction)
to column connection capillary 50 mounted on the injector.
•
with column heating:
Prepare column heating according to Section 2.9.2 of the
819/820 Instructions for Use and screw column connection
capillary 50 (see Fig. 20, 819/820 Instructions for Use) with a
compression fitting to injection valve 51.
•
With precolumn:
Install precolumn according to the supplied leaflet between
inlet of the separating column and the injection valve.
2 Rinse column
•
Place a beaker beneath the column outlet.
•
Start 818 IC Pump in «IC Net» with suitable flow (see leaflet of
the column) and rinse column for ca. 10 min with eluent.
•
Stop 818 IC Pump.
3 Connect column to suppressor module (Suppressor
connection 1)
•
Cut inlet capillary 64 (marked with "Eluent") of suppressor
module 70 to the required length. Use the 6.2126.080
Capillary cutter available as an option.
•
without column heating
Screw inlet capillary 64 on to the outlet end of separating
column 53 using a 6.2744.010 compression fitting.
•
with column heating
Connect capillary 52 at the outlet of the separating column
(see Fig. 20, 819/820 Instructions for Use) to the inlet
capillary 64 using the 6.2744.040 Coupling.
4 Fix column
•
without column heating
Insert one or two column holders 59 (6.2027.030, 6.2027.040
or 6.2027.050) in the mounting rails 58 and fasten separating
column in the column holder.
•
with column heating
Insert column heating according to Fig. 10 (819/820
Instructions for Use) into the Separation Center.