8.3 Connect filtration cell

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

73

7 Tubing connection injection valve – waste

Insert 6.2744.020 Coupling (from the 820 accessories) into

connection 13 provided for it in the 820 IC Separation Center.

Screw PTFE aspiration tubing 63 onto the 6.2744.020

Coupling inserted in connection 13 and lead it into the waste

bottle.

8 Rinse ultrafiltration cell

Each time that a new filtration membrane is inserted any air

that may still be contained in the filtration cell or tubing must

be removed. This is done by rinsing the cell and all the tubing

with e.g. ultrapure water:

Immerse aspiration tubing 16 (see Fig. 24) in the rinsing

solution (degassed ultrapure water).

Switch on the peristaltic pump in the «IC Net» software.

Rinse the filtration system with ultrapure water for approx. 5

min. Check that the same amount of solution emerges from

both the waste outlets in the waste bottle.

Check all tubing from the storage containers to the tubing

cassettes and filtration cell and up to the waste bottle for

leaks. If any leaks are found then the corresponding

connection must either be tightened up or replaced.

If any air bubbles remain in the dialysis cell then screw off

PEEK capillary 80 (filtrate) from outlet 71 and the filtration cell

and wait until the air bubbles have vanished. Then screw the

tubing back onto the filtration cell again.

8 Closing down the filtration cell

If the filtration cell is not to be used for a long time then it

must be rinsed with ultrapure water for approx. 10 min. The

inlet and outlet tubing is then screwed off from the filtration

cell and openings 71, 73 and 74 (see Fig. 22) are each

closed with a 6.2744.060 Blank stopper.

If the filtration cell is not to be used for a long time then the

filtration membrane must be removed and the cell cleaned,

see Section 9.2.8.1 and Fig. 22.

With 820.0X20 IC Separation Center equipped with two injection

valves it is possible to fill both sample loops from the same 833 IC

Liquid Handling Ultrafiltration Unit. In this case connection "1" at valve

A (sample loop outlet) must be connected to connection "2" of valve B

(sample loop inlet) using a 6.1831.040 PEEK capillary (15 cm).