3 Basic instrument
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
14
3 Basic instrument
3.1 Electrical
connection
Always switch off the 833 IC Liquid Handling Unit and the instrument
to which the Liquid Handling Unit is to be connected before you make
an electrical connection.
The 833 Liquid Handling Unit can be operated and remotely controlled
by a 761 Compact IC, an 819 IC Detector and in a modular system
via the 830 IC Interface.
3.1.1
Connecting to 761 Compact IC
Connect the remote connection of the 761 Compact IC with the
required remote connections 14 and 15 on the 833 IC Liquid Handling
Unit. The peristaltic pump of the Liquid Handling Unit is controlled via
remote connection Remote Pump 15, this must always be connected.
The connection Remote Actuator 14 is only required for the 833 IC
Liquid Handling Suppressor Unit and the 833 IC Liquid Handling
Sample Preparation Unit; it is used to switch the suppressor or
sample preparation module. The particular modules are connected
to the socket Module 13.
Fig. 4:
Connecting the 833 IC Liquid Handling Unit to 761 Compact IC
761
6.2143.210
816
833
761
6.2143.200
816
833
Cable
to
Module