beautypg.com

Metrohm 833 IC Liquid Handling Unit User Manual

Page 38

background image

4 833 IC Liquid Handling Pump Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

30

At the inlet end of the second piece of tubing 19 fasten a

suitably long piece of PTFE tubing 16 (6.1803.020) to

coupling 18 (6.2744.034) by using a compression fitting 17

(6.2744.010).

Immerse the other end of the aspiration tubing in a container

with rinsing solution (normally dist. H

2

O) and fix it in position.

Pull outlet capillary 67 of the suppressor module (marked

with "Waste") through opening 18 on the rear panel, lead it

into a sufficiently large waste bottle and fix it in position.

9 Start up the 833 Pump Unit

Switch on the 833 Pump Unit with mains switch 8.

Set the contact pressure for both cassettes: press contact

lever 5 until the solutions are just aspirated. Then press the

contact lever up by 1 further click to achieve the optimum

contact pressure.

Check all the tubing from the storage containers to the 833 IC

Liquid Handling Unit and the suppressor module through to

the waste bottles for leaks. If any liquid should be leaking

then the corresponding connection must be tightened up or

replaced.