9.2 Care and maintenance

830 IC Interface / Gebrauchsanweisung 8.830.1001

87

9.2.6.4

Adjusting the actuator step

It may be necessary to readjust the actuator step on the suppressor

module or sample preparation module in the following cases (see also

Section 9.2.6.3):

Suppressor/Sample preparation module does move to

next position but its functions are impaired.

After replacement of the actuator rotor (see Section

9.2.6.3)

The actuator step is adjusted by using the setting screw Step

Adjustment 12 on the rear panel of the 833 IC Liquid Handling Unit.

Proceed as follows:

1 Condition cation exchanger

If the actuator rotor has been cleaned or replaced the three

cation exchanger units must be conditioned before the

actuator rotor is moved to the next position for the first time.

Condition suppressor module

see Section 5.1.

Before the suppressor is switched to the next position for the

first time the actuator units should be rinsed with solution for

5 min.

Condition sample preparation module

See Section 6.1 and Section 6.2.

Before the sample preparation module is switched to the next

position for the first time the actuator units should be rinsed

with solution for 5 min.

2 Check position of actuator rotor

Switch the instrument off and then on again and activate the

suppressor in «IC Net» with .

By turning the suppressor module or using a mirror check the

position of the actuator rotor as shown in Fig. 27: in one of

the openings in the base of the suppressor module it should

be possible to see one of the three holes on the suppressor

rotor, at least partially.

Repeat the process 2-3 times; before each switch to the next

position with wait at least 10 s.

If the rotor stops in an invalid position then the step of the

actuator rotor must be readjusted (continue with item 3).