3 Basic instrument
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
16
3.1.2
Connecting to 819 IC Detector
Connect the remote connection of the 819 IC Detector to the necessary
remote connections 14 and 15 on the 833 IC Liquid Handling Unit. The
peristaltic pump of the Liquid Handling Unit is controlled via remote
connection Remote Pump 15, this must always be connected. The
connection Remote Actuator 14 is only required for the 833 IC Liquid
Handling Suppressor Unit and the 833 IC Liquid Handling Sample
Preparation Unit; it is used to switch the suppressor or sample
preparation module. The particular modules are connected to the
socket Module 13.
Fig. 5:
Connecting the 833 IC Liquid Handling Unit to 819 IC Detector
The following cables are available for these connections:
•
6.2143.200 Cable for one remote connection via
Remote line 1.
To connect a Liquid Handling Unit with a peristaltic pump, e.g.:
2.833.0010
IC Liquid Handling Pump Unit
2.833.0050
IC Liquid Handling Ultrafiltration Unit
•
6.2143.210 Cable for two remote connections via
Remote lines 1 and 2.
To connect a Liquid Handling Unit with a peristaltic pump and a
suppressor or sample preparation module, e.g.:
2.833.0020
IC Liquid Handling Suppressor Unit
2.833.0030
IC Liquid Handling Sample Preparation Unit
or
To connect two peristaltic pumps, e.g. for dialysis:
2.833.0010
IC Liquid Handling Pump Unit
2.833.0040
IC Liquid Handling Dialysis Unit
6.2143.210
816
833
6.2143.200
816
833
Cable
to
Module
819
819