7.7 Dialysis
procedure
For determinations of samples by using dialysis it is advisable to
proceed in the following sequence:
1 Prepare acceptor solution
•
To avoid interference from air bubbles in the acceptor
channel the ultrapure water used for the acceptor solution
should always be degassed for at least 10 min by vacuum,
N
2
or He.
2 Prepare sample
•
In order to avoid blocking the sample channel, samples
containing a large amount of suspended or solid particles
should always be centrifuged using a benchtop centrifuge
at 10'000 min
-1
for 5 min.
3 Start up IC system
•
Switch on all instruments.
•
Start the IC Net software and load/create the required system
(see «IC Net» Instructions for Use).
•
Condition the IC column (see 819/820 Instructions for Use).
4 Start up dialysis system
•
Immerse both pieces of aspiration tubing 16 in the acceptor
solution.
•
Start the peristaltic pumps in «IC Net» by setting the remote
line 1 (833 Pump Unit) and remote line 2 (833 Dialysis Unit) in
the window of the 819 Detector.
•
Rinse the dialysis system with acceptor solution for approx.
10 min and then switch off the pump again.
5 Calibration
•
Immerse sample aspiration tubing 16 in the standard
solution.
•
Start the determination in IC Net and wait until the standard
chromatogram has been evaluated and the next standard
can be measured.
6 Sample determination
•
Immerse sample aspiration tubing 16 in the sample solution.
•
Start the determination in IC Net and wait until the
chromatogram has been evaluated.