7.1 Flow diagram for dialysis

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

51

7.1

Flow diagram for dialysis

Sample dialysis with the 833 IC Liquid Handling Dialysis Unit comprises

the following four steps for each sample:

1 Rinsing

Valve

A: INJECT

Sample channel,

acceptor channel and

sample loop are rinsed

(for approx. 2 min).

2 Dialysis with «stopped flow»

Valve

A: INJECT

The sample is dialyzed

with stationary acceptor

solution (dialysis time

approx. 10 min).

3 Transfer to sample loop

Valve

A: FILL

The acceptor solution

enriched with ions from

the sample is

transferred to the

sample loop (transfer

time approx. 0.5 min).

4 Injection

Valve

A: INJECT


The acceptor solution

enriched with ions from

the sample is injected

into the IC system.

To perform dialysis with the 833 IC Liquid Handling Dialysis Unit two

peristaltic pumps are required. One pump (833 Dialysis Unit) is

delivering the acceptor solution, the other one (833 Pump Unit) is

A

Inject

833

Dialysis

Unit

833

Pump

Unit

A

Inject

833

Pump

Unit

833

Dialysis

Unit

Eluent

Colum

Waste

Sample

Acceptor

Waste

A

Inject

833

Dialysis

Unit

833

Pump

Unit

Fill

833

Dialysis

Unit

833

Pump

Unit

A

Eluent

Colum

Waste

Sample

Acceptor

Waste

Eluent

Colum

Waste

Sample

Acceptor

Waste

Eluent

Colum

Waste

Sample

Acceptor

Waste