beautypg.com

833 ic liquid handling suppressor unit – Metrohm 833 IC Liquid Handling Unit User Manual

Page 40

background image

5 833 IC Liquid Handling Suppressor Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

32

2 Connect suppressor module

820.0X20 IC Separation Center (2-channel system):

Remove plastic stopper from rear panel opening 18 of the

820 IC Separation Center and lead cable 26 of suppressor

module 24 out through one of the openings in the rear panel

of the IC Separation Center.

Connect cable 26 to connection 13 (“Module” of the 833 IC

Liquid Handling Unit (see Fig. 2).

3 Connect the column to the injector

Remove end caps from column 53

without column heating:

Screw inlet end of separating column 53

(note flow direction)

to column connection capillary 50 mounted on the injector.

with column heating:

Prepare column heating according to Section 2.9.2 of the

819/820 Instructions for Use and screw column connection

capillary 50 (see Fig. 20, 819/820 Instructions for Use) with a

compression fitting to injection valve 51.

With precolumn:

Install precolumn according to the supplied leaflet between

inlet of the separating column and the injection valve.

4 Rinse column

Place a beaker beneath the column outlet.

Start 818 IC Pumpe 818 in «IC Net» with suitable flow (see

leaflet of the column) and rinse column for ca. 10 min with

eluent.

Stop 818 IC Pump.

27 Suppressor inlet capillary for

eluent

28 Suppressor inlet capillary for

H

2

SO

4

29 Suppressor outlet capillary for

H

2

SO

4

30 Suppressor outlet capillary for

H

2

O

31 Suppressor inlet capillary for

H

2

O

32 Suppressor outlet capillary for

eluent

Fig. 10:

Suppressor module connections for 833 IC Liquid Handling
Suppressor Unit

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4

1

3

2

30

27

28

29

31

32