beautypg.com

2 care and maintenance, 1 instrument care, 2 maintenance by metrohm service – Metrohm 833 IC Liquid Handling Unit User Manual

Page 87: 3 shut down, Care and maintenance, Instrument care, Maintenance by metrohm service, Shut down

background image

9.2 Care and maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

79

9.2 Care

and

maintenance

9.2.1 Instrument

care

The 833 IC Liquid Handling Unit and its associated modules require a

reasonable amount of care. If the instrument becomes excessively dirty

this could affect the functioning and shorten the working life of the

basically robust mechanism and electronics.
Split chemicals and solvents should be cleaned up immediately. Above

all, the connection strip (the mains plug in particular) must be kept

clean.

Although this prevented to a large extent by constructive measures, if

aggressive media should penetrate the interior of the instrument then

the mains plug of the 833 IC Liquid Handling Unit should be pulled

out immediately in order to prevent massive damage to the

electronics. If such damage should occur please contact the Metrohm

service department.

The instrument must not be opened by untrained personnel. Please

observe the safety information given in Section 1.4.

9.2.2

Maintenance by Metrohm service

Servicing the 833 IC Liquid Handling Unit should be carried out within

the framework of an annual service visit by technicians from Metrohm. If

aggressive or corrosive chemicals are used then shorter service

intervals are necessary.
The Metrohm service department is always pleased to provide

competent advice about the servicing and maintenance of all Metrohm

instruments.

9.2.3 Shut

down

Suppressor module/Sample preparation module

If the IC system is not to be used for a long time then the whole IC

system (without column and suppressor module/sample preparation

module) must be rinsed salt-free with methanol/water (1:4) to prevent

eluent salts from crystallizing out and causing subsequent damage.

Rinsing is carried out by disconnecting the column and suppressor and

connecting the injector directly to the detector. Rinsing is carried out

with methanol/water (1:4) until the conductivity falls below 10

µ

S/cm.

Each of the three suppressor units is rinsed with ultrapure water for

approx. 5 min.