9 Troubleshooting - maintenance
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
84
5 Insert actuator rotor
•
Place actuator rotor 87 in actuator holder 88 so that the
tubing connections on the rear of the rotor fit into the
corresponding recesses inside the holder and one of the
three holes of the rotor can be seen from below in one of the
openings of the holder.
•
When the rotor has been correctly inserted its sealing surface
is approx. 4 mm inside the holder. If this is not the case then
the rotor must be brought into the correct position from below
by using a pointed object (e.g. screwdriver).
Fig. 26:
Actuator assembly
85 Sleeve nut
86 Connection piece with inlets and
outlets
6.2832.010 for suppressor module
or
6.2835.010 for sample preparation
module
87 Actuator rotor
6.2832.000 for suppressor module
or
6.2835.000 for sample preparation
module
6 Clean connection piece
•
Clean the sealing surface of connection piece 86 using a lint-
free cloth and ethanol.
7 Insert connection piece
•
Insert connection piece 86 in actuator holder 88 so that
connection "1" is at the top and the three connection piece
projections fit into the corresponding recesses in the holder.
•
Tighten up sleeve nut 85 on the thread of suppressor holder
88 by hand (do not use any tools).
87
86
85 88