7.5 Making the capillary connections

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

57

6 Assemble dialysis cell

Push the 6.2057.010 Cell holder into the free folding guide of

the 820 IC Separation Center and insert assembled dialysis
cell 42 in the holder.

Fig. 20

Inserting the dialysis cell in the holder

7.5

Making the capillary connections

7.5.1

Preparing the 820 IC Separation Center

For operation with the 833 IC Liquid Handling Dialysis Unit the 820 IC

Separation Center must be converted as follows (see Fig. 21):

1 Dismantle accessories

Pull the PTFE aspiration tube completely out of the

feedthrough 14 and screw it off from connection "1" of valve A

56.

Dismantle the PEEK capillary tubing from connection "2" of

valve A 56 to the feedthrough 13.

2 Make connections to suppressor (only for

2.820.0X30)

Install suppressor module of the 820 IC Separation Center

and the additional 833 IC Liquid Handling Pump Unit as

described in Section 4.