beautypg.com

Section 6.2 – Metrohm 833 IC Liquid Handling Unit User Manual

Page 51

background image

6.2 Connection to modular IC system for cation separation

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

43

7. The sample queue to be processed then contains the following

entries, for example:

System Ident

Vial Chrom.

No.

1

Start-Neutralisation.smt dummy 1

-

2

Neutralisation.smt Sample

1

1

1

3

Neutralisation.smt Sample

2

2

2

4

Neutralisation.smt Sample

3

3

3

5

Neutralisation.smt Sample

4

4

4

6

Neutralisation.smt Sample

5

5

5

7

End-Neutralisation.smt dummy 5

-

This example shows the processing of a sample series with 5 samples.

the 'Chrom. No.' column is only used for information.

6.2

Connection to modular IC system for cation
separation

A further application for the 833 IC Liquid Handling Sample Preparation

Unit is the separation of cations before injecting the sample onto the

separation column (e.g. separation of heavy metals). A system for the

determination of anions with chemical suppression is described, in

which the direct transfer of the sample to the sample preparation

module with subsequent transfer of the treated sample to the injection

valve is planned. This is why the necessary instrument configuration

corresponds to that of the Modular IC System 2 (MIC 2 Advanced:

anion system with chemical suppression), extended by the 833 IC

Liquid Handling Unit.
If you are not certain whether the capacity of the sample preparation

module is sufficient for your application then, for more strongly polluted

samples which require matrix elimination and enrichment, you should

use such a system as that described in Section 6.1.
The electrical connections of the system described above to the 766 IC

Sample Processor with complete control by the «IC Net 2.3» software

is described below.