Metrohm 833 IC Liquid Handling Unit User Manual
Page 23
3.1 Electrical connection
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
15
The following cables are available for these connections:
•
6.2143.200 Cable for one remote connection via
Remote line 1.
To connect a Liquid Handling Unit with a peristaltic pump, e.g.:
2.833.0010
IC Liquid Handling Pump Unit
2.833.0050
IC Liquid Handling Ultrafiltration Unit
•
6.2143.210 Cable for two remote connections via
Remote lines 1 and 2.
To connect a Liquid Handling Unit with a peristaltic pump and a
suppressor or sample preparation module, e.g.:
2.833.0020
IC Liquid Handling Suppressor Unit
2.833.0030
IC Liquid Handling Sample Preparation Unit
or
To connect two peristaltic pumps, e.g. for dialysis:
2.833.0010
IC Liquid Handling Pump Unit
2.833.0040
IC Liquid Handling Dialysis Unit
•
6.2143.220 Cable for three remote connections via
Remote lines 1, 2 and 4.
To connect up to three Liquid Handling Units, e.g. for dialysis with
suppression:
2x
2.833.0010
IC Liquid Handling Pump Unit
and 1x
2.833.0040
IC Liquid Handling Dialysis Unit
(three peristaltic pumps are switched on and off)
or
Ultrafiltration with suppression
1x
2.833.0020
IC Liquid Handling Suppressor Unit
and 1x
2.833.0050
IC Liquid Handling Ultrafiltration Unit
(two peristaltic pumps are switched on and off
and the sample preparation module is switched
on)
The remote line via which the control of the Liquid Handling Unit takes
place is determined by the wiring of the cable. The active remote line is
printed on the 833-end of the cable. This remote line must be set
accordingly in the control software «761 Compact IC» or «IC Net», see
Section 4.3.7 In the «761 Compact IC» Instructions for Use and Section
6.5.3 in the «IC Net» Instructions for Use.