3.2 Assembly the pump tubing
833 IC Liquid Handling Unit / Instructions for Use 8.833.1003
19
3.2
Assembly the pump tubing
Pump tubing is a consumable whose working life depends on the
contact pressure. This is why when the pump is switched off for a long
time, the tubing cassette should be lifted by fully loosening spring
lever
on the right-hand side (in this way the correctly set optimal
contact pressure is retained).
The 6.1826.0X0 Pump tubing consists of PVC or PP and must
therefore not be rinsed with solvents that contain acetone. In this case
you should either use different pump tubing or use a different pump
for rinsing.
Fig. 7:
Assembly the pump tubing
7 Spring lever
16 6.1803.0X0 Aspiration
tubeing/capillary
PTFE tubing/capillary, depends on
version, see Section 10.2 Standard
17 6.2744.010 PEEK compression
various tubing is available, depending
on the application, see Tab. 1.
20 Stopper
The color of the stopper indicates the
material and dimensions of the pump
tubing, see Tab. 1.
with tubing security device
22 6.2821.120 PEEK Filter unit
not contained in all versions, see
Section 10.2 Standard equipment
23 Discharge capillary
delivery side for connection to:
suppressor module
sample preparation module
dialysis cell
ultrafiltration cell
18
17
16 4
5
19
7
20 22
21
20 17
23
17
17
16