Metrohm 794 Basic Titrino User Manual
Page 91
3.2 Remote control commands
794 Basic Titrino
87
*Parameter
Tree part "Parameters for DET"
├
.
TitrPara
Titration parameters
│ ├
.MptDensity
Measuring point density
0...4...9 3.2.2.12.
│ ├
.MinIncr Minimum
increment
0...10.0...999.9 ditto
│ ├
.DosRate
Dispensing rate for increments
0.01...150.0, max. 3.2.2.13.
│ ├
.SignalDrift
Drift for meas. value acquisition
depends on meas.quant. 3.2.2.14.
│ ├
.UnitSigDrift
Unit of measured value drift
read only
ditto
│ ├
.EquTime Equilibrium
time
0...26...9999, OFF
ditto
│ ├
.StartV Start
volume
│ │ ├
.Type
Type of start volume
abs., rel., OFF 3.2.2.15.
│ │ ├
.V
Volume for absolute start volume
0...999.99 ditto
│ │ ├
.Factor
Factor for relative start volume
0...±999 999
ditto
│ │ ├
.Rate
Dispensing rate for start volume
0.01...150.0, max. ditto
│ ├
.Pause Waiting
time
0...999 999
3.2.2.16.
│ ├
.MeasInput Measuring
input
1, 2, diff.
3.2.2.17.
│ ├
.Ipol
Polarization
current
0...1...±127 ditto
│ ├
.Upol
Polarization voltage
0...400...±1270 ditto
│ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ ├
.Temp
Titration temperature
-170.0...25.0...500.0 3.2.2.18.
│
├
.
StopCond
Stop conditions
│ ├
.VStop
Stop volume
│ │ ├
.Type
Type of stop volume
abs., rel., OFF
3.2.2.19.
│ │ ├
.V
Volume for absolute stop volume
0...99.99...9999.99 ditto
│ │ ├
.Factor
Factor for relative stop volume
0...±999 999 ditto
│ ├
.MeasStop
Stop measured value pH, U, I
depends on meas.quant. 3.2.2.20.
│ ├
.UnitMStop
Unit of stop measured value
read only
ditto
│ ├
.EPStop
Stop after a number of EP's
1...9, OFF
3.2.2.21.
│ ├
.FillRate Filling
rate
0.01...150.0,
max. 3.2.2.22.
│
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.23.
│ ├
.MeanN
No. of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│
├
.
Evaluation
Evaluation
│ ├
.EPC
EP criterion
0...5...200 3.2.2.24.
│ ├
.Recognition EP
recognition
│ │ ├
.Select
Type of EP recognition
all,greatest,last,window,OFF ditto
│ │ ├
.Window
Window
│ │ │ ├
.1
up to 9 windows
│ │ │ │ ├
.LowLim
Lower limit window 1
depends on meas.quant.
ditto
│ │ │ │ ├
.UpLim
Upper limit window 1
depends on meas.quant.
ditto
│ ├
.FixEP
Fix endpoints
│ │ ├
.1
up to 9 fix EP's
│ │ │ ├
.Value
Measured value for fix EP1
depends on meas.quant.
ditto
│ ├
.pK pK or HNP evaluation
ON, OFF ditto
│
├
.
Presel
Preselections
│ ├
.IReq
Request of Id's after start
id1, id1&2, all, OFF 3.2.2.25.
│ ├
.Sreq
Request of smpl size after start
value, unit, all, OFF ditto
│ ├
.ActPulse
Output of a pulse
ON, OFF 3.2.2.26.