beautypg.com

Metrohm 794 Basic Titrino User Manual

Page 91

background image

3.2 Remote control commands

794 Basic Titrino

87

*Parameter

Tree part "Parameters for DET"

.

TitrPara

Titration parameters

│ ├

.MptDensity

Measuring point density

0...4...9 3.2.2.12.

│ ├

.MinIncr Minimum

increment

0...10.0...999.9 ditto

│ ├

.DosRate

Dispensing rate for increments

0.01...150.0, max. 3.2.2.13.

│ ├

.SignalDrift

Drift for meas. value acquisition

depends on meas.quant. 3.2.2.14.

│ ├

.UnitSigDrift

Unit of measured value drift

read only

ditto

│ ├

.EquTime Equilibrium

time

0...26...9999, OFF

ditto

│ ├

.StartV Start

volume

│ │ ├

.Type

Type of start volume

abs., rel., OFF 3.2.2.15.

│ │ ├

.V

Volume for absolute start volume

0...999.99 ditto

│ │ ├

.Factor

Factor for relative start volume

0...±999 999

ditto

│ │ ├

.Rate

Dispensing rate for start volume

0.01...150.0, max. ditto

│ ├

.Pause Waiting

time

0...999 999

3.2.2.16.

│ ├

.MeasInput Measuring

input

1, 2, diff.

3.2.2.17.

│ ├

.Ipol

Polarization

current

0...1...±127 ditto

│ ├

.Upol

Polarization voltage

0...400...±1270 ditto

│ ├

.PolElectrTest

Test for polarized electrodes

ON, OFF ditto

│ ├

.Temp

Titration temperature

-170.0...25.0...500.0 3.2.2.18.

.

StopCond

Stop conditions

│ ├

.VStop

Stop volume

│ │ ├

.Type

Type of stop volume

abs., rel., OFF

3.2.2.19.

│ │ ├

.V

Volume for absolute stop volume

0...99.99...9999.99 ditto

│ │ ├

.Factor

Factor for relative stop volume

0...±999 999 ditto

│ ├

.MeasStop

Stop measured value pH, U, I

depends on meas.quant. 3.2.2.20.

│ ├

.UnitMStop

Unit of stop measured value

read only

ditto

│ ├

.EPStop

Stop after a number of EP's

1...9, OFF

3.2.2.21.

│ ├

.FillRate Filling

rate

0.01...150.0,

max. 3.2.2.22.

.

Statistics

Statistics

│ ├

.Status

Status of statistics calculation

ON, OFF 3.2.2.23.

│ ├

.MeanN

No. of individual determinations

2...20 ditto

│ ├

.ResTab Result

table

│ │ ├

.Select

original,delete n,delete all

ditto

│ │ ├

.DelN

Deletion of individual results

1...20 ditto

.

Evaluation

Evaluation

│ ├

.EPC

EP criterion

0...5...200 3.2.2.24.

│ ├

.Recognition EP

recognition

│ │ ├

.Select

Type of EP recognition

all,greatest,last,window,OFF ditto

│ │ ├

.Window

Window

│ │ │ ├

.1

up to 9 windows

│ │ │ │ ├

.LowLim

Lower limit window 1

depends on meas.quant.

ditto

│ │ │ │ ├

.UpLim

Upper limit window 1

depends on meas.quant.

ditto

│ ├

.FixEP

Fix endpoints

│ │ ├

.1

up to 9 fix EP's

│ │ │ ├

.Value

Measured value for fix EP1

depends on meas.quant.

ditto

│ ├

.pK pK or HNP evaluation

ON, OFF ditto

.

Presel

Preselections

│ ├

.IReq

Request of Id's after start

id1, id1&2, all, OFF 3.2.2.25.

│ ├

.Sreq

Request of smpl size after start

value, unit, all, OFF ditto

│ ├

.ActPulse

Output of a pulse

ON, OFF 3.2.2.26.