Metrohm 794 Basic Titrino User Manual
Page 139
4.1 Error and special messages
794 Basic Titrino
135
stop V reached
The determination has been stopped as the stop volume has
been reached.
system error 3
The instrument adjustment data have been overwritten.
Exit:
message appears each time the instrument is switched on
until it has been readjusted (Metrohm service).
TIP terminated
TIP has been terminated.
wrong sample
With SET, with preset titration direction the first measured
value is outside the end point.
Error messages in connection with the data transfer
If neither a computer nor a printer is attached, the report output at the end of the titration
must be switched off:
Receive errors:
E36
Parity
Exit:
same on both instruments
E37
Stop
bit
Exit:
same on both instruments
E38
Overrun error. At least 1 character could not be read.
Exit:
E39
Overflow of the receive buffer of the Titrino (> 82
characters).
Exit:
Send errors:
E40
DSR=OFF
E41
DCD=OFF
E42
CTS=OFF
Handshake unsatisfactory for more than 1 s.
Exit:
to receive?
E43
The transmission of the Titrino has been interrupted
with XOFF for at least 3 s.
Exit: