Metrohm 794 Basic Titrino User Manual
Page 68
2.12. TIP, Titration procedure
794 Basic Titrino
64
Sequence of TIP
As there is no preset sequence of TIP, in what follows the procedure is illustrated by a
sequence that contains all available commands.
The start delay time is waited off.
The sample identifications and the sample size
are requested.
Submethods are processed according to their
parameters. They run to completion, including
calculations and data output (e.g. curves). The
determination data of submethods are
overwritten in the next sequence step of TIP.
Those values which are needed for higher-level
calculations must therefore be assigned to
temporary variables C7X.
The pause is waited off.
Outputs on the socket "Remote" can be set.
A message can be written in the display. The
sequence is held until it is continued manually
(with
TIP higher-level calculations are carried out.
Data output in TIP.
Attention: TIP contains no more determination
data, i.e. curves must be put out within the
submethods.
(Start delay)
(Request ident.)
(Request smpl size)
Submethod
Calculations
Data output
Set output
TIP calculations
Pause
Info
C7X
TIP data output