Metrohm 794 Basic Titrino User Manual
Page 107

3.2 Remote control commands
794 Basic Titrino
103
3.2.2.6. Mode.Def.SiloCalc.Assign.C24
RSX, EPX, CXX
Mode.Def.SiloCalc.Assign.C25
RSX, EPX, CXX
Mode.Def.SiloCalc.MatchId
id1, id1&2, all, OFF
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
Indication which sample identification(s) have to match so that
the results can be combined.
3.2.2.7. Mode.Def.ComVar.C30
RSX, MNX, EPX, CXX
Mode.Def.ComVar.C31
etc., up to .C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.58.
3.2.2.8. Mode.Def.Report.Assign
DET: full, short, mplist, curve, derive comb, scalc full, scalc srt,
calc, param, calib, ff
MET: full, short, mplist, curve, scalc full, scalc srt,
calc, param, calib, ff
SET, MEAS, CAL:full, short, scalc full, scalc srt, calc, param, calib, ff
TIP: full, short, scalc full, scalc srt, calc, param, ff
Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated with
";".
3.2.2.9. Mode.Def.Mean.1.Assign RS1, RSX, EPX, CXX
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a requirement
for statistics calculations. In addition, the statistics calculation must be
switched on, see 3.2.2.23. Rules for statistics calculations see page 55.
3.2.2.10. Mode.Def.TempVar.C70
RSX, EPX, CXX
etc. up to .C79
Assignment of temporary variables in a submethod for calculations in TIP.
3.2.2.11. Mode.CFmla
Mode.CFmla.1.Value
0...±999 999
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.60 and 3.2.2.61) and val-
ues of common variables (3.2.2.58) on the other hand are not stored with the
methods.
3.2.2.12. Mode.Parameter.TitrPara.MptDensity 0...4...9
Mode.Parameter.TitrPara.MinIncr 0...10.0...999.9