2.7 Parameters, key

794 Basic Titrino

41

Titration sequence of SET

After the start, the activate pulse is outputted.

The start delay time is waited off.

If conditioning is on, the sample solution is titrated until

the (first) EP is reached. The display shows then

drift OK 2.3 µl/min

or

SET pH conditioning

The vessel is now conditioned. The titration can be started

with .

The sample identifications and the sample size are re-

quested.

The temperature is measured if a T sensor is connected.

The

start volume is dispensed and pause 2 waited off.

Then the titration is performed to the first, then to the sec-

ond end point.

Calculations are carried out.

Data are outputted.

Conditioning is carried out.

(Activate pulse)

(Start delay)

(Start conditions)

(Request ident.)

(Request smpl size)

Titration:

1

st

end point

2

nd

end point

Calculations

Data output

(Reconditioning)

(Preconditioning)

(

(Activate pulse)

(Start delay)