3.2 Remote control commands
794 Basic Titrino
109
3.2.2.36. Mode.Parameter.Calibration.CalTemp -20.0...25.0...120.0
Parameters for CAL: Calibration temperature in °C. If a Pt 100 or Pt1000 is
connected, the temperature is measured.
3.2.2.37. Mode.Parameter.Calibration.Buffer.1.Value 0...7.00...±20.00
Mode.Parameter.Calibration.Buffer.2.Value 0...4.00...±20.00, OFF
etc. up to 9 buffers
Parameters for CAL: pH of buffers. The first buffer which is set to "OFF" de-
termines the number of buffers in the calibration.
3.2.2.38. Mode.Parameter.Calibration.SignalDrift 0.5...2...999, OFF
Mode.Parameter.Calibration.EquTime 0...110...9999, OFF
Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding cri-
terion is switched off. If both criterions are on OFF, the measured value is ac-
quired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 31. After it has been edited once, it
remains in force with the set value.
3.2.2.39. Mode.Parameter.Calibration.ElectrodeId
up to 8 ASCII char
Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.68.
3.2.2.40. Mode.Parameter.Calibration.SmplChanger ON, OFF
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.41. Mode.Parameter.Calibration.ActPulse first,
all, OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
163.
3.2.2.42. Mode.Parameter.Sequence.X.Select
method, pause, L4 output,
L6 output, info, OFF
Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.43.
3.2.2.43. Mode.Parameter.Sequence.X.Method Method
name
Mode.Parameter.Sequence.X.Pause
0...999 999, INF
Mode.Parameter.Sequence.X.L4Output
active, inactive,
pulse, OFF
Mode.Parameter.Sequence.X.L6Output
as for L4
Mode.Parameter.Sequence.X.Info
up to 16 ASCII characters