6.3. User methods

794 Basic Titrino

182

P2O5Fert TIP sequence

-

Start P2O5-1

-

Activaton of Dosimaten

-

Waiting time

-

Start P2O5-2

-

Content of P

2

O

5

in %

-

Sample size

-

Concentration of titrant

-

Equivalence weight of P

2

O

5

in g/mol

-

Excess of titrant in 1

st

titration in mL

-

Volume of titrant in 2

nd

titration in mL

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 10:06 0

TIP P2O5Fert

parameters

>sequence

1.method: P2O5-1

2.L6 output: pulse

3.pause 30 s

4.method: P2O5-2

>statistics

status: ON

mean n= 3

res.tab: original

>preselections

req.ident: OFF

req.smpl size: OFF

meas.mode: OFF

temperature 25.0 °C

------------

'fm

794 Titrino 01102 794.0010

date 2002-01-03 time 10:06 0

TIP P2O5Fert

>calculations

P2O5=(C70+C71)*C01*C02/C00;2;%

C00= 1.0

C01= 1

C02= 7.1

C70= 1.031

C71= 10.614

------------