4.1. Error and special messages

794 Basic Titrino

134

no method

The method required by the sample data from the silo memory

or in a TIP sequence is not available in the method memory.

Exit: .

no new com.var.

The common variable could not be assigned as the result or

the mean value could not be calculated. The old value remains

in force.

no new mean

No new mean value has been calculated as at least one

quantity stipulated for mean value calculations could not be

calculated.

no new silo result

No new silo result C24 or C25 could be stored as the assigned

quantity could not be calculated.

no new temp.var.

No new temporary variable C7X could be stored as the

assigned quantity could not be calculated.

no sequence

No sequence is defined in TIP.

Exit: and define sequence.

no titration data

No curve can be printed as no data are available.

not valid

A value is not available.

outside

The set fix EP is outside the measuring range.

overrange

The measuring range of ±2 V has been exceeded. Overrange

replaces the corresponding measured value (pH, U, I or

temperature).

Exit: Rectify error, or >MEAS/HOLD>.

same buffer

In the calibration sequence the voltage difference between the

first and second buffer is < 6 mV.

Exit: and change buffer or (abort

calibration).

save lines OFF

The function "save lines" is not active although a submethod of

TIP contains assignments to C24 or C25.

Exit: and switch on "save lines" under

DATA> key.

Warning: The results of this line will not be saved.

second TIP call

In TIP no further TIP can be called up as a submethod.

Exit: and define new sequence.

silo empty

The silo memory is switched in but empty and a titration has

been started. Corrective action: At least the first 1 silo line

before starting the first titration.

Exit: .

silo full

The silo memory is full up. Corrective action: If you have filled

less than 99 silo lines, you can create more space by deleting

old methods no longer needed. 1 silo line needs 18...120

bytes.

Exit: .

stop EP reached

A DET or MET titration was stopped as the stop criterion "stop

EP" was reached.

stop meas.val.reached

A DET or MET titration was stopped as the stop measuring

value pH, U or I was reached.

stop time reached

SET has been stopped as the stop time has been reached.