3.2 Remote control commands
794 Basic Titrino
111
UserMeth.List.1.Bytes read
only
UserMeth.List.1.Checksum read
only
for each method
List of the methods in the user method memory with the following
characteristics:
.Name:
Name of the method
.Mode: Mode
.Quantity: Measured
quantity
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.68.
3.2.2.48. Config.PeriphUnit.CharSet1
Epson, Seiko, Citizen, HP, IBM
Selection of the character set and the graphics control characters of the Ti-
trino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.49. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,
AND,Precisa
Config.PeriphUnit.Plot
U, dU/dt, V, dV/dt, U(rel), T
Selection of the balance type and the signal at the analog output.
3.2.2.50. Config.Aux.Language english, deutsch, francais, espanol,
italiano, portugese, svenska
Selection of the dialog language.
3.2.2.51. Config.Aux.Set $G
Config.Aux.Set.Date YYYY-MM-DD
Config.Aux.Set.Time HH:MM
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.
3.2.2.52. Config.Aux.RunNo
0...9999
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
3.2.2.53. Config.Aux.AutoStart 1...9999, OFF
Number of automatic, internal starts.
3.2.2.54. Config.Aux.StartDelay
0...999 999
Start delay time in s. During this time, the data of the preceding determination
are retained.