3.2. Remote control commands
794 Basic Titrino
114
Silo lines are not deleted when they are worked off. Assigned results are
stored as C24 and C25. "Save lines" can only be set to "ON" if the silo is
completely empty. Delete the silo, see 3.2.2.64.
3.2.2.67. Info.Report $G
Info.Report.Select
configuration, parameters, smpl data,
statistics, silo, calib, C-fmla, def,
user method, full, short, mplist,
curve, deriv, comb, scalc full, scalc
srt, calc, all, ff
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running deter-
mination.
parameters: Parameter report of the current method. During a running deter-
mination only "live"-parameters are accessible.
smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo:
Contents of the silo memory.
calib:
Calibration data of the measuring input in the current method.
C-fmla:
Contents of the
def:
Contents of the
user method: Contents of the method memory.
full:
Full result report of the last completed determination.
short:
Short result report of the last completed determination.
mplist:
Measuring point list of the running determination.
curve:
Titration curve of the last determination.
derive:
1st derive of titration curve of the last determination (with DET).
comb:
1st derive combined with the titration curve of the last determina-
tion (with DET).
scalc full: Full report of the silo calculations.
scalc srt: Short report of the silo calculations.
calc:
Calculation report of the current method.
all: All
reports.
ff:
Form feed on printer.
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.68. Info.CalibrationData $G
Info.CalibrationData.Inp1.pHas -20.00...7.00...+20.00
Info.CalibrationData.Inp1.Slope -9.999...1.000...+9.999
Info.CalibrationData.Inp1.Temp -170.0...25.0...+500.0
Info.CalibrationData.Inp1.Date read
only
Info.CalibrationData.Inp1.ElectrodeId read
only
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.39.