4 error messages, troubleshooting, 1 error and special messages – Metrohm 794 Basic Titrino User Manual
Page 137
![background image](/manuals/586355/137/background.png)
4.1 Error and special messages
794 Basic Titrino
133
4 Error messages, Troubleshooting
Data transfer inoperative See measures on page 131.
4.1 Error and special messages
XXX bytes missing
XXX bytes missing. For the storage of a method or a silo line
XXX bytes are missing or there is insufficient RAM for a TIP
sequence.
Remedy:
fewer silo lines.
check electrode
With polarized electrodes. There is a break or short circuit.
Possible causes and rectification of the fault:
- the electrode is not plugged in ⇒ plug it in
- the electrode is not immersed in the solution ⇒ immerse it
- the electrode is defective ⇒ use new electrode.
- the electrode cable is defective ⇒ use new cable.
The electrode test can be switched off under the
key.
Exit: Rectify fault or
check exchange unit
The Exchange Unit is not mounted (properly).
Exit: Mount Exchange Unit (properly) so that the coupling
engages or
check T-sensor
No temperature sensor is attached (with MEAS T or if the
temperature monitoring is active).
Exit: Connect Pt100 or Pt1000 or
data set reevaluation
Message if Titrino is reevaluating downloaded measuring point
lists.
division by zero
The result could not be calculated as a divisor in the formula
was equal to zero.
Exit: Enter appropriate value.
EP overflow
In a DET or MET titration 9 EP's or more were found. The first 9
EP's will be listed.
Remedy: Recalculation of data with higher EP criterion.
manual stop
The determination has been manually stopped.
meas.pt list overflow
Maximum 500 measured points can be stored.
Exit: Use start criteria or select larger time interval.
missing EP
An EP needed for calculation in a formula is missing.
missing fix EP
A fix EP needed for calculation in a formula is missing.
no.EP not corresponding
In DET or MET titrations, the number of EP's actually found
does not match the set windows: Exactly 1 EP per window has
not been found.
no EP set
In SET, no EP has been set.
Exit:
no meas.quantity
In TIP a manual measurement (
performed without defining a measuring quantity.
Exit: