Metrohm 794 Basic Titrino User Manual
Page 119

3.2 Remote control commands
794 Basic Titrino
115
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.69. Info.Checksums $G
Info.Checksums.MPList read
only
Info.Checksums.ActualMethod read
only
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList: Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.70. Info.DetermData $G
Info.DetermData.Write ON,
OFF
Info.DetermData.MPList.1.Attribute read
only/read+write
Info.DetermData.ExV read
only/read+write
Info.DetermData.MPList.1.X read
only/read+write
Info.DetermData.MPList.1.Y read
only/read+write
for every measuring point
Determination data in hexadecimal format. A measuring point list is available
in mode DET, MET, SET, and MEAS.
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
3.2.2.71. Info.TitrResults.RS.1.Value read
only
etc., up to .9
Info.TitrResults.EP.1.V read
only
Info.TitrResults.EP.1.Meas read
only
etc., up to .2
Info.TitrResults.Var.C40 read
only/read+write
etc., up to .C47
.RS: Values of the calculated results.
.EP: Endpoints with DET, MET, SET:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in pH "5.12", mV (with U and Ipol) "-
241" or µA (with Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.95.