3 "blank, 3 user methods – Metrohm 794 Basic Titrino User Manual
Page 171

6.3 User methods
794 Basic Titrino
167
6.3.3 "Blank"
Treat the "blank sample" in the same way as
you treat your real samples. Can be used e.g.
for the methods Sapon.No. and Perox.No,
see pages 172 and 176.
The blank value is stored as common variable
C30 and can therefore be used directly by
subsequent methods.
Reference:
METROHM Application Bulletin Nr. 210:
– Result as reagent consumption in mL
– Division factor if multiple quantities of solvent are
used for the blank determination.
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 09:38 0
DET U Blank
parameters
>titration parameters
meas.pt.density 4
min.incr. 10.0 µl
titr.rate max. ml/min
signal drift 50 mV/min
equilibr.time 26 s
start V: OFF
pause 0 s
meas.input: 1
temperature 25.0 °C
>stop conditions
stop V: abs.
stop V 30 ml
stop U OFF mV
stop EP 9
filling rate max. ml/min
>statistics
status: ON
mean n= 3
res.tab: original
>evaluation
EPC 5
EP recognition: all
fix EP1 at U OFF mV
pK/HNP: OFF
>preselections
req.ident: OFF
req.smpl size: OFF
activate pulse: OFF
------------
'fm
794 Titrino 01102 794.0010
date 2002-01-03 time 09:38 0
DET U Blank
>calculations
Blank=EP1/C01;3;ml
C01= 1
------------