3.2 Remote control commands
794 Basic Titrino
121
Setup.Lock.UserMeth.Recall ON, OFF
Setup.Lock.UserMeth.Store ON, OFF
Setup.Lock.UserMeth.Delete ON, OFF
Setup.Lock.Display ON, OFF
ON means disable the corresponding function:
.Keyboard: Disable all keys of the Titrino
.Config:
Disable the
.Parameter: Disable the
.SmplData: Disable the
.UserMeth.Recall: Disable "recall" in
.UserMeth.Store: Disable "store" in
.UserMeth.Delete: Disable "delete" in
.Display: Disable the display, i.e. it will not be written to by the device pro-
gram of the Titrino and can be operated from the computer.
3.2.2.92. Setup.Mode.StartWait ON, OFF
Setup.Mode.FinWait ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP
(holding point after AutoInfo !".T.GC").
.FinWait: Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").
3.2.2.93. Setup.SendMeas.SendStatus ON, OFF
Setup.SendMeas.Interval 0.08...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
3.2.2.95 and 3.2.2.96) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.95 and 3.2.2.96. The
inputted value is rounded off to a multiple of 0.08. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the inter-
face and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.
The automatic transmission is switched on/off with 'SendStatus'.
3.2.2.94. Setup.SendMeas.Select Assembly,
Titrator
Selection of the unit of which the measured values should be sent (3.2.2.95
or 3.2.2.96).
3.2.2.95. Setup.SendMeas.Assembly.CyclNo ON, OFF
Setup.SendMeas.Assembly.V ON, OFF
Setup.SendMeas.Assembly.Meas ON, OFF
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.93):
.CyclNo: Cycle number of the potential measurement. Together with the
cycle time (3.2.2.82), a time frame can be set up.