Metrohm 794 Basic Titrino User Manual
Page 23

2.4
Tutorial
794 Basic Titrino
19
If the titration runs too slow or too fast, you can change the control parameters during the
titration. If you wish to titrate faster, change the following parameters:
• dynamics: lower value, the control range becomes smaller (Attention: possibility of
titrating over the set EP)
• max.rate: bigger value
• min.rate: bigger value
You will find further details of the control parameters on page 42.
After the titration, end point volume and pH value at the end point are displayed.
For the evaluation enter:
the formula to calculate the m value
(key
RS1=EP1*C01*C02
with an accuracy of
RS1 decimal places
2
the unit
RS1 unit:
mmol/L
and the calculation constants
C01
1 (concentration of titrant ×10)
(key
C02
4 (factor for the sample size 100 mL/25 mL)
If you have a printer connected, select the automatic titration report (key
port:full". You can print out the results of the previous titration with keys
If the previous titration has run to your satisfaction, you can start thinking about perform-
ing statistics calculations. Have you already added a new sample to the titration vessel? If
you are no longer certain, you can find out immediately with
Rapid measurement between titrations
Press
measurement with a second
Statistics calculations
Now switch on the statistics calculations. Press
on. Duplicate determinations are now performed.
• Perform 2 titrations.