beautypg.com

8 "sapon.no – Metrohm 794 Basic Titrino User Manual

Page 176

background image

6.3. User methods

794 Basic Titrino

172

6.3.8 "Sapon.No"

Determination of the saponification number of

edible oils and fats.

Electrode:

6.0229.100 Solvotrode, at measuring input 1.

6.2320.000 Elektrolyte c(TEA-Br) = 0.4 mol/L

in ethylene glycol.

Titrant:

c(HCl) = 0.5 mol/L

Sample:

Weigh in app. 2 g of sample. Add 25 mL

c(KOH) = 0.5 mol/L in ethanol and allow to

boil lightly for at least 30 miutes. Rinse the

content into a beaker and titrate the excess

KOH with HCl.

Determine a blank sample the same way.

Remarks:

Change the result unit e.g. to mg KOH.

Reference:

Metrohm Application Bulletin Nr. 141

– Saponification number in mg KOH per g sample

– Sample size in g

– Molar mass of KOH * normality of titrant

(56.10*0.5)

– Consumption of "blank sample" in mL

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 09:50 0

DET U Sapon.No

parameters

>titration parameters

meas.pt.density 4

min.incr. 10.0 µl

titr.rate max. ml/min

signal drift 50 mV/min

equilibr.time 26 s

start V: OFF

pause 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 99.99 ml

stop U OFF mV

stop EP 1

filling rate max. ml/min

>statistics

status: OFF

>evaluation

EPC 5

EP recognition: all

fix EP1 at U OFF mV

pK/HNP: OFF

>preselections

req.ident: OFF

req.smpl size: value

activate pulse: OFF

------------

'fm

794 Titrino 01102 794.0010

date 2002-01-03 time 09:50 0

DET U Sapon.No

>calculations

Sapon.No=(C30-EP1)*C01/C00;2;mg/g

C00= 1.0

C01= 28.05

C30= 0.0

------------