Metrohm 794 Basic Titrino User Manual
Page 33

2.7 Parameters, key
794 Basic Titrino
29
stop EP 9
**titr.
filling rate max. ml/min
**titr.
Stop after a number of EP's have been found (1...9,
OFF)
"OFF" means that the criterion is not monitored.
Filling rate (0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.
>evaluation
EPC 5
EP recognition: all
EP evaluation/recognition
See page 32ff.
Equivalence point criterion (input range depends on
mode.
For DET:
0...200
For MET:
pH: 0.10...9.99
U, Ipol:
1...999 mV
Upol: 0.1...99.9
µA)
Threshold for the size of the jump, see page 34.
Recognition of EP's which fulfill the EP criterion. (all,
greatest, last, window, OFF)
Selection of equivalence point recognition:
"all":
All equivalence points are recognized.
"greatest": Only the greatest (steepest) equivalence
point is recognized.
"last":
Only the last equivalence point is
recognized.