2.7 Parameters, key

794 Basic Titrino

29

stop EP 9

**titr.




filling rate max. ml/min

**titr.

Stop after a number of EP's have been found (1...9,

OFF)

sets "OFF".

"OFF" means that the criterion is not monitored.

Filling rate (0.01...150 mL/min, max.)

sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.


>evaluation


EPC 5









EP recognition: all






EP evaluation/recognition

See page 32ff.

Equivalence point criterion (input range depends on

mode.

For DET:

0...200

For MET:

pH: 0.10...9.99

U, Ipol:

1...999 mV

Upol: 0.1...99.9

µA)

Threshold for the size of the jump, see page 34.

Recognition of EP's which fulfill the EP criterion. (all,

greatest, last, window, OFF)

Selection of equivalence point recognition:

"all":

All equivalence points are recognized.

"greatest": Only the greatest (steepest) equivalence

point is recognized.

"last":

Only the last equivalence point is

recognized.