beautypg.com

Metrohm 794 Basic Titrino User Manual

Page 123

background image

3.2 Remote control commands

794 Basic Titrino

119

3.2.2.82. Info.Assembly.CycleTime read

only

Info.Assembly.ExV read

only

Inquiries regarding basic variables of the assembly: Cycle time in s, volume of

the active Exchange Unit in mL.

3.2.2.83. Assembly.Bur.Rates.Forward.Selected digital, analog

Assembly.Bur.Rates.Forward.Digital 0...150,

max.

Assembly.Bur.Rates.Reverse.Selected digital, analog

Assembly.Bur.Rates.Reverse.Digital 0...150,

max.

Expel and aspirating rate.

Digital or analog control. With digital control, the inputted value applies (in

mL/min). "max." means maximum possible rate with the Exchange Unit in

current use.

Analog means rate control with the analog potentiometer on Titrino.

3.2.2.84. Assembly.Bur.Fill

$G, $H, $C

$G starts the 'FILL' mode of the buret function.

3.2.2.85. Assembly.Bur.ModeDis

$G, $S, $H, $C

Assembly.Bur.ModeDis.Selected volume, time

Assembly.Bur.ModeDis.V 0.0001...0.1...9999

Assembly.Bur.ModeDis.Time 0.25...1...86400

Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF

Assembly.Bur.ModeDis.AutoFill ON, OFF

Dispensing mode with parameters. The dispensing mode can only be started

and stopped via the RS Control. During a running dosification, no method can

be started at the Titrino.

.Selected: Dispensing of volume increments or during a preset time.

.Volume, .Time: Size of the volume increments or entry of time.

.VStop:

Limit volume for the dispensing.

.AutoFill: ON means automatic filling after every dispensing.

3.2.2.86. Assembly.Meas.Status ON, OFF

Assembly.Meas.MeasInput

1, 2, Diff., Ipol, Upol, Temp

Assembly.Meas.Ipol ±127...1...+127

Assembly.Meas.Upol ±1270...400...+1270

Measurement in assembly. The measuring function can only be started via RS

Control. When the measuring function is switched on, no method can be

started at the Titrino.

.Input:

Selection of the potentiometric measuring input 1, 2, diff., polar-

ized electrodes or temperature.

.Ipol:

Polarization current in µA.

.Upol:

Polarization potential in mV, entry in steps of 10 mV.

3.2.2.87. Assembly.Outputs.AutoEOD ON, OFF