2.7 Parameters, key

794 Basic Titrino

31

Titration sequence of DET and MET

After the start, the activate pulse is outputted.

The start delay time is waited off.

The sample identifications and the sample size are re-

quested.

The start volume is dispensed (no meas.value acquisition)

and the pause is waited off.

During the titration the volume increments are dispensed

and after each increment a measured value is acquired.

Meas.values are either acquired drift controlled ("equilib-

rium titration") or a after a fixed waiting time. If you have

not intentionally changed the equilibration time, it is calcu-

lated according to the formula:

equilibr.time (in s)

5

0.01

Drift

150

+

+

=

The criterion (drift or equilibration time) which is first met

applies. This avoids "infinite" titrations. If the parameter

"signal drift" is set "OFF", the measured values are ac-

quired after a fixed equilibration time.

The titration is terminated according to the first criterion

which is met.

Evaluations and calculations are carried out.

Data are outputted.

(Activate pulse)

(Start delay)

(Start conditions)

(Request ident.)

(Request smpl size)

Titration:

Dispense increments

Acquire meas.values

Stop conditions

Calculations

Data output